EasyManua.ls Logo

Steris InnoWave Unity - 12.5 Decontamination Program

Steris InnoWave Unity
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Observed problem:
Check Chemical appears on the display.
Possible causes and solutions:
l Make sure that there is sufficient chemical required by the wash program and
replace the bottle if there is insufficient chemical.
l Make sure that the chemical pickup is fully-inserted into the chemical bottle.
l Make sure that the chemical pickup is not kinked or restricted.
l Touch OK on the display after making sure that the chemical bottle and chemical
pickup are in a satisfactory condition. The wash program resumes.
Observed problem:
Instruments not clean after wash cycle.
Possible causes and solutions:
l Make sure that only a recommended chemical is used. Make sure that the
chemical is used according to manufacturer instruction.
l Instruments may be excessively soiled or have been left soiled for more than 2
hours.
l The basket may be overloaded. Do not load more than the recommended
number of instruments for cleaning.
l Make sure that at least one luer hose is connected to each flush manifold on the
basket.
l The supply power may not be sufficient for correct equipment performance.
Make sure that the supply power meets the specified requirements.
96 TROUBLESHOOTING
InnoWave Unity & Unity 20 Operator Manual

Table of Contents

Related product manuals