10.
Rinsing
On the left is the
water temperature
(0°Cto+60°C, or+32°Fto+140°F).
In the center is the
duration
in minutes
(0to20).
On the right is the
number of times
the
rinsing stage is performed (0to3).
If either duration or number of times is set
to zero, rinsing is not performed.
11.
Final rinse
On the left is the
water temperature
(0°Cto+60°C, or+32°Fto+140°F).
On the right is the
duration
in minutes
(0to20).
If either value is set to zero, final rinse is
not performed.
12.
Save Changes
If an unlocked wash program has been
edited, the operator is prompted to save
the changes, or cancel the edit.
If no changes have been made, only the
cancel option is enabled, and is used to
return to the supervisor menu main
selections.
Touch OK to save changes made to an
unlocked program, or touch Cancel to
discard any changes made to a program
or exit viewing a locked prgram.
82 THE SUPERVISOR MENU
InnoWave Unity & Unity 20 Operator Manual