6.
Pre-wash
On the left is the
duration
of the pre-wash
stage in minutes (0to20).
On the right is the
number of times
the
pre-wash stage is performed (0to3).
If either value is zero, pre-wash is not
performed.
7.
Degassing
In the center is the
duration
of the
degassing stage in minutes (0to10).
If the value is set to zero, degassing is not
performed.
8.
Washing
On the left is the quantity of
chemical
dispensed (0to2000ml, or
0to67.6fl.oz).
In the center is the
water temperature
(0°Cto+60°C, or +32°Fto+140°F).
On the right is the
duration
in minutes
(0to40).
9.
Rewash
In the center is the
number of times
the
tank fluids will be used for washing
(0to10).
If value is set to zero, rewash is not
performed and tank fluids are drained
after every washing stage.