1-2
764333-675
WARNING SLIPPING HAZARD:
WARNING SHOCK AND BURN HAZARD:
WARNING POTENTIAL BURN HAZARD:
The SYSTEM 1E Liquid Chemical Sterilant Processing System is not protected against water spillage
or spray. Take precautions in the use of liquid during cleaning or insertion of devices.
Connect the SYSTEM 1E Liquid Chemical Sterilant Processing System to a hospital-grade, GFI
electrical receptacle only, to prevent electrical shock hazard.
SYSTEM 1E Processor performance has been validated and optimized with components defined by
STERIS in the Operator Manual for the SYSTEM 1E Processor. Use of pre-filters that do not meet
STERIS's specifications can lead to premature failure of the MaxPure Filter inside the processor.
STERIS’s warranty will not apply to damage or early life failure of the MaxPure Filter resulting from
the use of pre-filters that do not meet STERIS’s specifications. DO NOT USE components that do not
meet STERIS’s specifications.
The UV System unit may be hot to the touch!
The UV System is not user serviceable. All installation, maintenance work and service must be
carried out by an authorized STERIS Service Representative only. Use only STERIS-approved
replacement lamps, quartz sleeves and monitors.
UV chamber may be under pressure. Before attempting to remove end plates, monitor or quartz
sleeve, depressurize and drain chamber.
Disconnect all utilities to SYSTEM 1E Liquid Chemical Sterilant Processing System before servicing.
Do not service the SYSTEM 1E Liquid Chemical Sterilant Processing System unless all utilities have
been properly locked out. Always follow OSHA Lockout-Tagout and electrical safety-related work
practice standards.
Lid assembly is heavy. Always support lid assembly when gas spring cylinder is removed.
To avoid slippery floor conditions, immediately wipe up any spilled liquids in the SYSTEM 1E Liquid
Chemical Sterilant Processing System
loading and unloading areas.
Disconnect all utilities to SYSTEM 1E Liquid Chemical Sterilant Processing System before servicing.
Do not install processor unless all utilities have been properly locked out. Always follow all locally
mandated Lockout-Tagout and electrical safety-related work practice standards.
Possible high temperature. Before attempting to dismantle, make sure UV chamber is in a cool state.
WARNING PERSONAL INJURY AND/OR EQUIPMENT DAMAGE
HAZARD: (CONTINUED)
!
!
!
!
!
!
!
!
!
!
!