EasyManuals Logo

Steris SYSTEM 1E Service Manual

Steris SYSTEM 1E
179 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
1-3
764333-675
WARNING LIQUID CHEMICAL STERILANT VERIFICATION HAZARD:
Failure to thoroughly clean devices prior to processing may result in an ineffective liquid chemical
sterilization process.
Use the Quick Connect specifically designed for the device being processed.
Do not use S40™ Sterilant Concentrate after the expiration date indicated on the label.
Do not substitute any other product for S40 Sterilant Concentrate.
Use of a blocked or damaged aspirator may result in ineffective liquid chemical sterilization.
Always verify that the sterilant container is empty at the end of the cycle.
Devices are not liquid chemically sterilized and/or adequately rinsed when a liquid chemical sterilant
processing cycle is cancelled.
Do not adjust inlet water pressure regulator from factory-set value of 50 psig.
Do not adjust Ultra Violet (UV) system hose lengths.
!
!
!
!
!
!
!
!
!
CAUTION POSSIBLE EQUIPMENT DAMAGE HAZARD:
Do not energize “h”, control screen designation for heater element (HTR1), unless chamber is full of
water; element failure will occur.
Do not energize “s2”, control screen designation for seal release valve (SOL2), with water in the
chamber; spill will occur.
Do not force lid to close.
After a cycle is cancelled, if fluid remains in the chamber, call STERIS.
When replacing the control board, use care when picking up the sheet metal/board assembly, because
the display and printer cables are still connected under the board.
In the Service Mode, do not energize Heater unless chamber is full of water; element failure will occur.
In the Service Mode, do not energize Seal Relief Valve/Compressor with water in the chamber; spill
will occur.
To avoid jamming paper in printer, do not force paper into printer mechanism and do not attempt to
feed uneven, tattered, wrinkled paper.
!
!
!
!
!
!
!
!

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Steris SYSTEM 1E and is the answer not in the manual?

Steris SYSTEM 1E Specifications

General IconGeneral
BrandSteris
ModelSYSTEM 1E
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals