3. Protect electronics.
Place an absorbent towel over electronic enclo-
sures and printer paper to protect from water
drops.
CAUTION POSSIBLE EQUIPMENT
DAMAGE HAZARD: Water spills on the
microcomputer housing and/or the printer
or printer paper may result in damage to the
SYSTEM 1E Processor or create a paper jam.
4. Remove housing cap of filter.
Unscrew housing cap of filter by turning it coun-
terclockwise.
5. Remove filter.
Pull up on metal ring on top of the filter while
twisting to remove filter.
6. Drain water from filter.
While holding filter over filter housing, permit
water to drain from filter for 20-30 seconds.
7. Check replacement filter type.
Examine replacement filter to verify it is exactly
the same as the one being replaced (looks alike,
same catalog number).
If it is the same filter, proceed to Step 8.
!