8. Replace with new MaxPure filter.
Cut plastic wrap on seam near filter connection
collar with O-ring.
CAUTION POSSIBLE EQUIPMENT
DAMAGE HAZARD: Do not remove
MaxPure filter from plastic bag until fully
installed.
Dip end of filter in water to wet O-ring.
9. Insert filter. Align water filter core with copper tube
in central housing of the MaxPure Filter. Slide filter
out of plastic bag with O-ring down. Once MaxPure
filter is in housing, press on top of filter to verify it
is sealed (top edge of MaxPure Filter housing).
Remove plastic bag.
10. Replace housing cap of filter. Line up threads and
turn clockwise until snug.
11. Remove absorbent towels.
12. Close control panel.
13. Plug in SYSTEM 1E Liquid Chemical Sterilant
Processing System.
14. Press "Change Values" button until message:
"MAXPURE FILTER REPLACED?" appears.
Use arrow button to select "YES" and press "Save
Values" button.
15. Initiate a Diagnostic Cycle.
A successful Diagnostic Cycle must be
completed.
!