WWaatteerr tteemmppeerraattuurree
Control setting range 39.2° - 104.0° F 4.0° - 40° C
Control accuracy ±0.3° C (4.0° - 40.0° C)
Display measurement accuracy ±0.2° C (4.0° - 40.0° C)
Display / resolution setting 0.1° C
Default setting 104.0° F 40.0° C
PPaattiieenntt tteemmppeerraattuurree
Control setting range 89.6° - 100.4° F 32.0° - 38.0° C
Control accuracy ±0.1° C (32° - 38°C)
Measurement accuracy ±0.3° C (25.0° - 45.0° C)
±0.4° C (0° C - 24.9° C, 45.1° C - 50° C)
Display / resolution setting 0.1° C
Display range 32.0° - 122° F 0.0° - 50° C
Default setting 98.6° F 37.0° C
CCoonnttrroolllleerr
Heater capacity, maximum 500 watts
Circulating fluid Sterile distilled water or distilled water that has been passed through a filter less than
or equal to 0.22 microns with this product
Battery 9V Lithium
Alarm tone range 75 - 85 dBA per standard IEC 60601-1-8
Water flow rate in each hose port Typical 1.2 lpm
Refrigerant type R134a
Power cord length 14 to 15 feet 4.2 - 4.5 meters
Clinical thermometer Direct mode
Equipment Class Class I
Rate for continuous operation
NNoottee -- The controller takes approximately 9 minutes to heat from 23.0±2° C (73.4° F) to 37.0° C (98.6° F) when not
connected to a patient. Time will vary when connected to a patient.
Stryker reserves the right to change specifications without notice.
For more information about thermal transfer devices, cables, or probes, see the manufacturer’s instructions for use.
EN 8 8001-109-001 Rev C.0