SSeettuupp
Unpack the cartons and check all items. Make sure that the product is free from visual damage before putting into service.
CCAAUUTTIIOONN
Shock Hazard - Improper handling of the power cord may damage the power cord and cause potential shock hazards. If
damage has occurred to the power cord, immediately remove the temperature management system from service to
avoid the risk of serious injury or death. Contact the appropriate maintenance personnel.
Take special precautions regarding electromagnetic compatibility (EMC) when using medical electrical equipment like
AAllttrriixx. Install and place AAllttrriixx into service according to the EMC information located in the EMC section of this manual.
Portable and mobile RF communications equipment can affect the function of AAllttrriixx.
Shock Hazard. If the internal electrical components are exposed, because the side panel or cover are compromised,
remove the product from use.
Always make sure that the product reaches room temperature before you setup or operate the product.
Before first use, disinfect the internal water circuit.
IInnssppeeccttiinngg
Before you place the product into service, make sure that the controller works.
1. Visually inspect the product for any signs of shipping damage.
2. Plug the product into a properly grounded, hospital grade wall receptacle. Make sure that the power indicator illuminates
on the operator control panel.
3. Before first use,
Disinfect the internal water circuit and hoses every 14 days
(page 39).
SSeelleeccttiinngg aa llaanngguuaaggee
The AAllttrriixx controller has several language choices. English is the default language.
To select a language when in standby mode:
1. Tap the Settings button, to show the Select Language screen.
a. Tap Next, if you are in therapy mode.
2. Tap More to view other languages.
3. Select a language. Tap on the Increase or Decrease buttons or tap the language, to highlight a language of your choice.
4. Tap Confirm.
NNoottee -- If you do not touch the screen for three minutes, the LCD will return to the previous menu.
TTeessttiinngg vviissuuaall aanndd aauuddiibbllee aallaarrmmss
Before placing the product into service, make sure that the visual and audible alarms are functioning.
1. Tap the Settings button.
2. Tap the Back button.
3. Tap the Visual/Audible icon.
EN 18 8001-109-001 Rev C.0