TTrroouubblleesshhoooottiinngg
PPrroobblleemm PPoossssiibbllee ccaauussee AAccttiioonn RReemmoovvee ffrroomm uussee
Controller will not turn on Power cord is not plugged
into a properly grounded
hospital grade wall outlet
Insert the plug fully into the
properly grounded hospital
grade wall outlet
If the product does not turn
on after trying a different
outlet
Damaged power cord or
plug
Visually make sure that the
power cord is not damaged
If damaged, RFU
Controller user interface
blackout
Power outage If the Stand-by button is
solid green, visually inspect
the LCD for damage
If damaged, RFU
Product alarming, user
interface blackout
Power outage If the Stand-by button is
yellow and flashing, visually
inspect the LCD for damage
If damaged, RFU
Temperature probe Not responding, does not
connect, temperature
outside of range
Replace temperature probe
Check connections
If damaged, RFU
Controller will not heat Reservoir is empty Fill reservoir, tap Confirm
that sterile distilled water
has been added, restart
If filling the reservoir does
not resolve
Controller will not cool Reservoir is empty Fill reservoir, tap Confirm
that sterile distilled water
has been added, restart
If filling the reservoir does
not resolve
Thermal transfer device not
filling with water or ports not
detecting flow
Locking ring on Clik-Tite
connector is not snapped
into place
Check the CClliikk--TTiittee
connection
Replace the cable or
thermal transfer device
Thermal transfer device
may be to high, lower the
bed level
Thermal transfer device
may be folded, lay flat to
make sure the water flows
Not applicable
Quick disconnect not seated
properly
Secure the thermal transfer
device connection to the
controller
Replace the cable or
thermal transfer device
Not applicable
Water level alarm Water level too low Fill reservoir Not applicable
EN 48 8001-109-001 Rev C.0