PPrroobblleemm PPoossssiibbllee ccaauussee AAccttiioonn RReemmoovvee ffrroomm uussee
Patient temperature Out of range Check the placement of the
probe
Not applicable
Patient temperature output
(PTO)
External device output
displays high value >45° C
when input is out of range
(As a result of one of the
following: patient probe
disconnected, controller in
standby / sleep mode,
patient temperature outside
the range of 25° - 45°C).
To resume calibration:
Disconnect the external
device from the reusable
adapter cable
Tap the Help button to
display the alarm
screens
Find the “Temperature
Output Alarm” screen
Tap Confirm to restart
calibration
Wait until the “Monitor” is
solid
Connect the external
device to the reusable
adapter cable
Not applicable
8001-109-001 Rev C.0 49 EN