PPrroobblleemm PPoossssiibbllee ccaauussee AAccttiioonn RReemmoovvee ffrroomm uussee
Patient temperature Out of range Check the placement of the
probe
Not applicable
Patient temperature output
(PTO)
External device output
displays high value >45° C
when input is out of range
(As a result of one of the
following: patient probe
disconnected, controller in
standby / sleep mode,
patient temperature outside
the range of 25° - 45°C).
To resume calibration:
• Disconnect the external
device from the reusable
adapter cable
• Tap the Help button to
display the alarm
screens
• Find the “Temperature
Output Alarm” screen
• Tap Confirm to restart
calibration
• Wait until the “Monitor” is
solid
• Connect the external
device to the reusable
adapter cable
Not applicable
8001-109-001 Rev C.0 49 EN