IIccoonn NNaammee
Total duration
Visual and audible tests
Target patient or water temperature
Medium: patient temperature increases at a rate of 4.0° C in 12 hours (0.33° C/ hour).
Maximum: water temperature approaches water target as fast as possible
Minimum: patient temperature increases at a rate of 4.0° C in 24 hours (0.17° C/hour).
Set Custom: patient temperature increases at a customized temperature and time period the
operator selects. The temperature increases 0.05° C/hour to 0.5° C/hour
Maximum: water temperature approaches water target as fast as possible
Medium: water is cooled to target, with a max of 15.0° C difference between the patient and the
water temperature
Minimum: water is cooled to target, with a max of 10.0° C difference between the patient and
the water temperature
PPrroodduucctt aallaarrmmss
Audible alarms work in conjunction with the display.
AAllaarrmm pprriioorriittyy aanndd ddeessccrriippttiioonn
PPrriioorriittyy aallaarrmm AAuuddiibbllee rreemmiinnddeerr IIccoonn ffllaasshheess
Medium Repetitive burst of 3 beeps upon first occurrence
and every 25 seconds until the condition is
resolved.
If you pause the audio, when the timer expires,
the audio indication will resume.
When a medium priority alarms, the icon will
breathe yellow to indicate there is an alarm. The
yellow icon will continue to breathe until the
alarm is resolved.
Low Single burst of 2 beeps upon first occurrence. The icon for the low priority alarm is solid yellow.
The solid yellow icon will remain until the alarm
condition is resolved.
Audio pause The audio pause button breathes as a reminder
and resets all of the active alarms.
The audible alarm will resume as soon as a new
alarm is raised or the last Medium audio paused
timer has expired.
You can press the audio pause button at any time
during the alarm condition, as many times as
necessary.
Pausing the alarm will not stop the yellow icon
from breathing or being solid yellow.
EN 14 8001-109-001 Rev C.0