IIccoonn,, yyeellllooww NNaammee AAllaarrmm pprriioorriittyy
aanndd ddeellaayy
MMeessssaaggee TThheerraappyy
iinntteerrrruupptteedd
CChheecckk
Water
temperature
deviation
Medium, 5 minute
delay
Water
temperature is
±0.8° C (1.4° F)
outside of target
temperature
No Temporary condition
upon startup,
addition of thermal
transfer device, or
addition of water
No water Medium, 60
second delay
No water Yes Check for leaks
Add a minimum of 2
liters of water
No water flow Medium, 20
second delay
No flow detected Yes Check for leaks and
obstructions at
connections, hoses,
and thermal transfer
devices
Check water flow Medium, 10
minute delay
Reduced flow
detected
No Tap Confirm, if the
water port was
removed intentionally
Check for leaks and
obstructions at
connections, hoses,
and thermal transfer
devices
Check patient
probe (A or B)
Medium Abnormal change
in patient
temperature
Yes Check probe
condition, location,
and connections
Probe or adapter
malfunction (A or
B)
Medium, 30
second delay
No temperature
signal detected.
Yes Check probe or
adapter cable
condition, location,
and connections
Adapter cable
disconnected (A
or B)
Medium, 30
second delay
Adapter cable not
detected
Yes Reinsert the adapter
cable. If damaged,
replace the adapter
cable
Patient
temperature
deviation
Medium Patient
temperature is
more than ± 1.0°
C from target
temperature
(Only will appear
after the initial
target is reached.)
No Check patient
condition, placement
of thermal transfer
devices, and all
connections
Patient
temperature
deviation
Low Patient
temperature is
more than ± 0.5°
C from target
temperature
(Only will appear
after the initial
target is reached.)
No Check patient
condition, placement
of thermal transfer
devices, and all
connections
8001-109-001 Rev C.0 15 EN