IIccoonn,, yyeellllooww NNaammee AAllaarrmm pprriioorriittyy
aanndd ddeellaayy
MMeessssaaggee TThheerraappyy
iinntteerrrruupptteedd
CChheecckk
Normothermia
deviation
Low Patient
temperature is
outside of 36.0° C
(96.8° F) to 38.0°
C (100.4° F)
No Check patient
condition, placement
of thermal transfer
devices, and all
connections
Therapy pause Medium Therapy is
currently paused
Yes To resume, press
and hold the play /
pause for 2 seconds
Battery low Low Battery is low No Maintenance is
recommended. If
battery is not
replaced, the product
may not function on
the next startup.
Patient
temperature
output (A or B)
Low
Patient
temperature
output is
inaccurate on the
external device,
or outside
supported range
No Check output
adapter cable
connection. Tap
Confirm to reactivate
the output port.
Remove from use
(RFU)
Medium The system has
powered off due
to a malfunction
Yes Remove the product
from use
immediately. Notify
the appropriate
personnel.
Power loss Medium Not applicable Yes Check power cord
connection
NNoottee
If any of the alarm conditions persist, call maintenance.
If page indicators appear on the alarm screen, there are multiple active alarms. The highest alarm is displayed. Tap Next
or Back to view the active alarms.
CCoonnttaacctt iinnffoorrmmaattiioonn
Contact Stryker Customer Service or Technical Support at: 1-800-327-0770.
Stryker Medical
3800 E. Centre Avenue
Portage, MI 49002
USA
NNoottee -- The user and/or the patient should report any serious product-related incident to both the manufacturer and the
Competent authority of the European Member State where the user and/or patient is established.
To view your operations or maintenance manual online, see https://techweb.stryker.com/.
Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support.
Include the serial number in all written communication.
EN 16 8001-109-001 Rev C.0

Table of Contents

Other manuals for Stryker altrix

Related product manuals