CCAAUUTTIIOONN
• Avoid the use of materials of good thermal conductivity, such as water, gel, or similar substances, with the AAllttrriixx system
not powered on. This can decrease the temperature of the body of a patient.
• Do not apply thermal transfer devices to patients with ischemic limbs. This may result in harm to the patient.
• Do not use this product if the patient has a transdermal medication (patch) as this can result in increased drug delivery.
• Do not use three or more adult Mul-T-Blanket products at the same time to avoid the risk of water overflow when you
power off the controller.
• Always pre-fill the thermal transfer devices with sterile distilled water before you apply it to the patients. This is to reduce
the risk of pressure ulcers.
• Always clamp the hoses when disconnecting the thermal transfer devices.
To connect or disconnect the CClliikk--TTiittee® connectors (Figure 12) to the insulated hoses.
FFiigguurree 1122 –– CClliikk--TTiittee
To connect or disconnect the Colder style (Figure 13) to the insulated hoses.
FFiigguurree 1133 –– CCoollddeerr ssttyyllee ccoonnnneeccttoorrss
To close or open hose clamps (Figure 14).
Always clamp the hoses before disconnecting. See
Draining the thermal transfer devices
(page 31).
FFiigguurree 1144 –– HHoossee ccllaammppss
To connect the AAllttrriixx®®TTeemmppeerraattuurree MMaannaaggeemmeenntt WWrraapp™™ to the AAllttrriixx®®TTeemmppeerraattuurree MMaannaaggeemmeenntt HHoossee™™ see
Figure 15.
Press the button (A) on the hose to disconnect from the wrap.
FFiigguurree 1155 –– AAllttrriixx TTeemmppeerraattuurree MMaannaaggeemmeenntt WWrraapp ccoonnnneeccttoorrss
The AAllttrriixx®®TTeemmppeerraattuurree MMaannaaggeemmeenntt WWrraappss have an internal valve and do not need a clamp to stop the water flow.