AAllaarrmm ccoonnddiittiioonnss
The alarm rank establishes the order of presentation of the alarm on-screen message. The D in the table indicates the
alarm is deactivated during that mode. Maintenance and RFU modes are always in the deactivated mode and are not listed
in the table.
This product maintains the individual alarm status for all alarms as defined below.
Alarm condition present
Visual indicator state
On screen message
Audible indicator state
Current timer for audio pause activation
Alarm rank per therapy mode
AAllaarrmm ((mmeedd)) SSttaanndd--bbyy AAuuttoo AAuuttoo
PPaauusseedd
MMaannuuaall MMaannuuaall
PPaauusseedd
MMoonniittoorr
Remove from use 0 0 0 0 0 0
Power loss D 1 1 1 1 1
Check patient probe D 7 7 11 11 4
Patient probe malfunction D 6 6 10 10 3
Probe disconnected D 5 5 9 9 2
Patient temperature deviation D 9 9 D D D
Water temperature deviation D D D 7 D D
Check water flow (all ports) D D D 6 D D
No water flow alarm (all ports) D 4 D 4 D D
No water D 2 D 2 D D
Therapy paused timed out D D 3 D 3 D
AAllaarrmm ((llooww)) SSttaanndd--bbyy AAuuttoo AAuuttoo
PPaauusseedd
MMaannuuaall MMaannuuaall
PPaauusseedd
MMoonniittoorr
Normothermia deviation D D D D D 5
Patient temperature deviation D 12 12 D D D
Power backup level 1 19 19 14 14 6
Patient output deviation D 22 22 17 17 7
NNoottee
If more than one alarm is active at the same time, the product maintains the active state for the individual alarm including
the audio pause timer. The screen alarms are displayed with the highest level alarm first with a page toggle to allow the
operator to scroll to the subsequent alarms.
The Paused in Auto Paused and Manual Paused refer to the Therapy Pause state.
EN 52 8001-109-001 Rev C.0