9. Discard the disposable thermal transfer devices based on your local waste management protocol.
a. Discard the disposable thermal transfer devices based on your local waste management protocol, if applicable.
DDrraaiinniinngg wwaatteerr ffrroomm tthhee rreesseerrvvooiirr
To drain the water from the reservoir:
1. See
Removing and replacing the reservoir
(page 25).
2. Dispose of the water per hospital protocol.
3. Replace the reservoir.
NNoottee -- Make sure that the reservoir is dry before you store the product.
DDrraaiinniinngg wwaatteerr ffrroomm tthhee ccoonnttrroolllleerr aanndd hhoosseess
Make sure that the controller and all components are dry before you store the product. Make sure to drain the hoses before
you store them.
1. Place the controller over a floor drain.
2. Remove the reservoir and pull up on the controller drain plug (A) to open the drain (Figure 21).
A
FFiigguurree 2211 DDrraaiinn pplluugg
3. Connect a hose to each port.
a. If you have Colder style connector hoses, attach the service tool adapter hose (8001-999-017).
b. If you have CClliikk--TTiittee hoses, make sure that the connectors and clamps are open (Figure 22).
EN 32 8001-109-001 Rev C.0