FFiigguurree 2288 –– CClliikk--TTiittee hhoossee eennddss aarree cclloosseedd aanndd ccllaammppss aarree ooppeenn
6. To fully drain the hoses, raise all the hoses (Figure 29) above the connection ports on the controller.
NNoottee -- For best performance, hang the hoses to keep them raised. Do not lower the hoses until you have completed the
disinfection and rinsing process.
FFiigguurree 2299 –– RRaaiissee tthhee hhoosseess
7. Allow the controller and hoses to drain for a minimum of two minutes.
8. Push down on the drain plug to close the drain.
DDiissiinnffeeccttiinngg tthhee iinntteerrnnaall wwaatteerr cciirrccuuiitt aanndd hhoosseess
1. Use personal protection equipment as recommended by the BBrruuCClleeaann TTbbCC disinfectant manufacturer’s instructions for
use.
2. Put 2 BBrruuCClleeaann TTbbCC tablets into the reservoir.
3. Using appropriate measuring equipment, fill the empty reservoir with 1 gallon (3.8 L) of sterile distilled water.
NNoottee -- AAllwwaayyss aallllooww tthhee ddiissiinnffeeccttaanntt ttaabblleettss ttoo ccoommpplleetteellyy ddiissssoollvvee bbeeffoorree ssttaarrttiinngg tthhee 2200 mmiinnuuttee ddiissiinnffeeccttiioonn
ccyyccllee..
4. Place the reservoir into the controller.
5. Disconnect the bottom hose from the bottom right port (Figure 30).
8001-109-001 Rev C.0 41 EN