DDiissiinnffeecctt tthhee iinntteerrnnaall wwaatteerr cciirrccuuiitt aanndd hhoosseess eevveerryy 1144 ddaayyss
Use the BBrruuCClleeaann TTbbCC disinfectant tablets manufactured by BBrruulliinn (EPA registration number 71847-2-106) before first
use, at least every 14 days, and before storage. BBrruuCClleeaann TTbbCC has been validated for internal water circuit disinfection.
Make sure that you follow the disinfectant manufacturer’s guidelines to avoid the risk of injury. Failure to follow the
disinfectant’s instructions may void your warranty.
CCAAUUTTIIOONN
• Always use sterile distilled water or distilled water that has been passed through a filter less than or equal to 0.22
microns with this product.
• Do not disinfect the internal water system with a thermal transfer device attached as this may cause a leak.
• Do not use bleach or any other cleaning or disinfectant agents for internal circuits. This could result in damage to the
product. Only use approved disinfectant tablets.
• Always drain the product before disinfecting the internal water circuit. Failure to drain the product may reduce the
effectiveness of the disinfection process.
NNoottee -- Disinfection of the AAllttrriixx internal water system was validated using
M. mucogenicum
.
TToooollss RReeqquuiirreedd::
• 2 gallons (7.6 L) of sterile distilled water or distilled water that has been passed through a filter less than or equal to 0.22
microns
• Personal protection equipment (PPE) as recommended by the disinfectant manufacturer’s instructions
• Soft, lint free cloth (2 or more)
• 2 BBrruuCClleeaann TTbbCC tablets (active ingredient is sodium dichloroisocynaturate (NaDCC); 1874 ppm disinfectant solution
when dissolved in one gallon of water)
• Service tool adapter hose (8001-999-017) for CCoollddeerr style connector hoses
• Floor drain
See
Product illustration
(page 9) for clarification of product component names and locations.
DDrraaiinniinngg tthhee iinntteerrnnaall wwaatteerr cciirrccuuiitt aanndd hhoosseess ffoorr ddiissiinnffeeccttiioonn
1. Unplug the power cord from the wall outlet.
2. Place the controller over a floor drain.
NNoottee -- For best results, the floor drain should be within reach of a wall outlet to power on the controller.
3. To drain the controller, pull up on the controller drain plug (A) to open the drain (Figure 25). Leave the drain open.
8001-109-001 Rev C.0 39 EN