GGuuiiddaannccee aanndd mmaannuuffaaccttuurreerr’’ss ddeeccllaarraattiioonn -- eelleeccttrroommaaggnneettiicc iimmmmuunniittyy
Portable and mobile RF
communications equipment
should be used no closer to
any part of the AAllttrriixx
system, including cables,
than the recommended
separation distance
calculated from the equation
appropriate for the
frequency of the transmitter.
Recommended separation
distance
D=(0.35) (√
P
)
80 MHz to 800 MHz
D=(0.70) (√
P
)
800 MHz to 2.7 GHz
where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site survey,
a
should be less than the
compliance level in each
frequency range.
b
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
Conducted RF
IEC 61000-4-6
3 Vrms
6 Vrms in ISM bands
150 kHz to 80 MHz
3 V
6 V in ISM bands
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
8001-109-001 Rev C.0 59 EN