EasyManua.ls Logo

Stryker ENDOSCOPY Standard - Testing Overpressure Alarm; Testing Blockage Alarm; 2-10 Testing Gas Heater

Stryker ENDOSCOPY Standard
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
US-11
2-8 Testing Overpressure Alarm
1. Attach a tube and an air-filled syringe to insufflation tube
connection (Fig. 2-5-1).
2. Use the syringe to slowly generate a pressure of 15 mm Hg.
3. Slowly increase pressure to 21 mm Hg and press the Start/
Stop key.
4. An alarm is sounded after approx. 5 seconds and
Overpressure is displayed.
5. Select gas flow level 2.
6. The bleeder (relief) system is activated after approx. 5
seconds (time and pressure value can be changed in the
user menu).
7. The bleeder system relieves the pressure; alarm and
Overpressure warning symbol are turned off.
8. Press the Start/Stop key and remove test tools and
resources.
Adjustment
Replace low-pressure unit.
2-9 Testing Blockage Alarm
1. Connect insufflation tube and close off the end of the tube.
2. Press the Start/Stop key.
3. An alarm is sounded after approx. 5 seconds and
Blockage is displayed. The acoustic warning signal can be
deactivated in the user menu.
4. Open the insufflation tube connection; the alarm is turned
off.
5. Press the Start/Stop key.
Adjustment
Return device to factory.
2-10 Testing Gas Heater
1. Connect a heater tube to insufflation tube and gas heater
connection.
2. The display depicts HEATING. The tube is being pre-
warmed.
3. Unplug the gas heater. The display HEATING disappears.
4. Remove heater tube.
5. Turn off device.
Adjustment
Replace HPL board or cable to HPL.
2 Function Test

Related product manuals