US-35
7 Technical Data
Power Supply 100 -240 V~
Main Fuse T 3,15 A
USA 3,15 A (slow blow), UL-recocnized
220-240 V~
Main Fuse T 3,15 A, UL-recocnized
Connection for potential equalization
Frequency 50-60 Hz
Max. Power Consumtion 130 /150*W
Max. Current 100 V: 1250/1400 *mA
240 V: 540/620* mA
Protection Class I, Typ BF, IP41
Dimensions Width x Height x Depth
273 x 145 x 360 [mm]
10,75x5,7x14,2 [inch]
Weight Approx. 7 kg
Operation Conditions 10-40 °C / 50-104 °F
30-75 % rel. air humedity
Storage and Transportation Directions -40 - +70 °C /40 - +158°F
10-85 % rel. air humedity
85-100% rel. air humedity (14 days)
Manufactured and Tested acc. to EN 60601-1 / IEC 601-1
EMV EN 60601-1-2 / IEC 601-1-2
CE 93/42/ EWG
Insufflation medium Medical CO
2
Maximun output pressure 55 mm Hg
Maximun gas supply pressure 80 bar/1160 PSI
Minimum gas supply pressure 5 bar/73,3 PSI
Measurement range of gas supply 0-50 bar/0-725 PSI
Maximun gas flow refer to device data plate located on rear of the device
Pressure range 1-30 mm Hg
Accuracy of pressure measurement ±5 %
Accuracy of gas flow measurement ±5 %
Accuracy of volume measuremant ±10 %
Accuracy of gas supply measurement ±10 %
Conections (optional) Video S-VHS IN/OUT
Video FBAS IN/OUT
Video RGB IN/OUT
RS232 Service Interface
* with reusable haeting tube