www.stryker.com 72-0908E OM FL28C REV C 19
Return To Table of Contents
FOOT BOARD CONTROL PANEL GUIDE
A: Press to raise/lower the Fowler.
B: Press to raise/lower the Gatch.
Note
The bed is equipped with the Auto Contour positioning. This function partially raises the Gatch as the Fowler is raised.
It prevents the patient from slipping toward the foot end of the bed. The function is operational by default. To deactivate
it: press the Gatch lockout control (D3), and the padlock icon will light up. Note that the Gatch functions on the siderail
control panels will not operate longer while the lockout is activated.
C: Press to raise/lower the bed.
D1: Press to lockout all the controls related to the mattress support sections and bed height. The padlock icon will
light up. The total lockout control deactivates the foot end and siderail controls.
Note
The total lockout will not affect the operation of the nurse call (optional), Scale (optional) or Bed Exit (optional) systems.
D2, D3: Press to lockout the inner and outer siderail control panel Fowler (D2) and Gatch (D3) functions. The
corresponding padlock icon will light up.
Note
The foot end controls are not affected by lockouts D2 and D3. In the event of a power failure, the lockout settings will
be saved.
F1: Press to obtain the desired Trendelenburg (head down, foot up) position. To return the sleep surface to the
horizontal position, simply use the bed Hi-Lo controls to either raise or lower the sleep surface to its limits.
F2: Press to obtain the desired reverse Trendelenburg (head up, foot down) position. To return the sleep surface
to the horizontal position, simply use the bed Hi-Lo controls to either raise or lower the sleep surface to its limits.
D1
D2
D3
C
F1
BA
F2
SCALE CONTROLS
(OPTIONAL), PAGE 20
BED EXIT CONTROLS
(OPTIONAL), PAGE 24
Figure 2.14A
Operation Guide