www.stryker.com 72-0908E OM FL28C REV C 33
Return To Table of Contents
Patient Helper - FA64148L
The overhead trapeze patient helper provides a support that allows the patient to change position in bed. The system
features a fixed lower section, a mobile upper section and a trapeze. When in use, the upper section is centered over
the patient. When not in use, the upper section can rotated to either side of the bed for storing.
WARNING
The scale and bed exit systems must be adjusted if this accessory is added when either system is in operation. Refer to
the “Add / Remove Equipment” procedure (page 22).
INSTALLATION
Tools Required:
• None
Procedure:
1. Insert the lower section posts (A) into the slots on the head
end of the bed. Ensure the posts seat completely into the
slots
2. Insert the hitch pins (B) into the holes (C) located on the ends
of each post.
3. Place the nylon washer (D) onto the pivot pin (E) of the lower
section.
4. Pull the lock pin (G) and install the upper section onto the
pivot pin (E) of the lower section.
5. Verify the lock pin (F) on the upper section is seated in an
adjustment hole.
OPERATION
WARNING
• The overhead trapeze is designed to help the patient change
position in bed. The trapeze must not be used as a support to
enter or exit the bed. Injuries may result due to improper use
of the overhead trapeze. Medical advice may be necessary
to determine if the use of an overhead trapeze is adequate
for the patient’s condition or treatment given.
• The maximum load capacity of the overhead trapeze is 150
lb (68 kg).
1. To position the trapeze over the patient, pull the lock pin (F)
and pivot the upper section towards the center of the bed.
Release the lock pin and ensure it is properly engaged.
2. To adjust the trapeze (H) height, shorten or lengthen the
strap (J). If you must detach the strap to free the trapeze,
ensure the strap is properly reinserted in the fastener.
A
E
D
F
G
K
H
J
C
C
B
B