Guidance and Manufacturer’s declaration - Electromagnetic Immunity
The GoBed
®
II MedSurg Bed, Model FL28C is suitable for use in the electromagnetic environment specified below.
The customer or the user of the GoBed
®
II MedSurg Bed, Model FL28C should assure that it is used in such an
environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic
Environment Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+6 kV contact
+8 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrostatic fast
Transient/burst
IEC 61000-4-4
+2 kV for power
supply lines
+1 kV for input/
output lines
+2 kV for power
supply lines
+1 kV for input/
output lines
Main power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+8 kV differential mode
+2 kV common mode
+8 kV differential mode
+2 kV common mode
Main power quality is that of
a typical commercial and/or
hospital environment.
Voltage dips, voltage variations
and short interruptions on
power supply input lines
IEC 61000-4-11
<5%Ut (95% dip in Ut) for
0.5 cycle
40%Ut (60% dip in Ut) for
5 cycles
70%Ut (30% dip in Ut) for
25 cycles.
<5% Ut (>95% dip in Ut)
for 5 sec.
<5%Ut (95% dip in Ut)
for 0.5 cycle
40%Ut (60% dip in Ut)
for 5 cycles
70%Ut (30% dip in Ut)
for 25 cycles.
<5% Ut (>95% dip in Ut)
for 5 sec.
Main power quality should be
that of a typical commercial
and/or hospital environment.
If the user of the GoBed
®
II
MedSurg Bed, Model FL28C
requires continued operation
during power main interruptions,
it is recommended that the
device be powered from an
uninterrupted power supply or
a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
and/or hospital environment.
Note: U
T
is the a.c. mains voltage prior to applications of the test level.
GOBED
®
II MEDSURG BED, MODEL FL28C
EMC Information