BED EXIT SYSTEM WITH MULTI-ZONE CONTROL (OPTIONAL) (CONTINUED)
Arming the System
Following the zeroing of the system, the Bed Exit system may now be activated:
• Install the patient in the bed. Ensure the patient lies in the bed in its usual position before activating the system.
• Activate the Bed Exit by pressing ARM/DISARM (K) and select the desired zone.
• By default, the least restrictive zone will be selected and its LED (left LED) will come on. To change the
zone, press ZONE (L) repeatedly until the desired zone LED comes on.
• To deactivate the Bed Exit after an alarm or to simply turn it off, press ARM/DISARM (K). The zone LED will turn
off.
Adding or Removing Equipment with the System Armed
Changing equipment on a bed with the Bed Exit system armed requires the user to follow either procedure described
below:
• If the bed is equipped with the scale system, follow the “Adding or Removing Equipment when a Patient is on the
Bed” on page 22.
-or-
• If the bed is not equipped with the scale system, remove the patient from the bed; change the equipment and zero
the system (see ¨Zeroing the System¨ on page 24). The Bed Exit system may be activated once the patient returns
to the bed.