26
Required Supplies
AIM IC-GREEN® (NDC # 17478-801-02) includes the required ICG supplies.
Other items required to use AIM IC-GREEN shall be provided by the user.
Note: In the United States, the L10 LED Light Source with AIM Technology
and SafeLight Fiber Optic Cable are indicated for use only with AIM IC-GREEN.
For users outside the U.S., refer to Stryker user manual P27009.
Item Description
AIM IC-GREEN Includes six 25mg vials of ICG powder and
six 10mL ampules of Sterile Water for
Injection, USP
Sterile normal saline Approximately 10 –12mL following each ICG
injection
Syringe
(for ICG preparation)
Minimum volume of syringe: 10mL
Syringes
(for each ICG solution
injection)
Minimum volume of syringes: 3 or 5mL
Syringes
(for each saline ush
after ICG injection)
Minimum volume of syringes: 12mL
ICG Preparation
Prepare the ICG solution following the steps below. The instructions yield a
2.5mg/mL solution.
1. Open the Sterile Water for Injection, USP ampule.
2. Draw 10mL of the solvent into a syringe.
3. Remove the ip cap o the vial containing 25mg of the ICG powder.
4. Inject the 10mL of Sterile Water for Injection, USP into the ICG vial
through the stopper to dissolve the powder.
5. Gently shake the ICG vial to mix the solution. Ensure all ICG powder is
dissolved in the solution.
6. Inspect the reconstituted solution for precipitation. If a precipitate is
present, discard the solution and prepare a new vial as described above.
ICG Dosage
The typical dose administered is 2.5mg.
The total dose should be less than 2 mg/kg per patient.