Replacing the Fuses
To avoid the risk of re, use only fuses of the value specied on
the fuse label located on the rear panel of the console.
1. Unplug the power cord from the wall outlet and remove the cord from
the light source console.
2. Unlatch the fuse holder above the AC inlet and remove it. (You may need
to press the tab on the fuse holder with a slender screwdriver to release
the latch.)
3. Replace the fuse with the same value and rating.
4. Reinstall the fuse holder until the tab snaps in place.
Storage
Never store the light source in a non-ventilated, humid environment. This can
damage the delicate electronics in the device.
Expected Service Life
The L10 AIM Light Source has an expected service life of 2000hours (2years
based on 1000cases per year).
Disposal
This product contains electrical waste or electronic equipment.
It must not be disposed of as unsorted municipal waste and
must be collected separately in accordance with applicable
national or institutional related policies relating to obsolete
electronic equipment.
Dispose of the light source according to local laws and hospital practices.
Refer to the diagram below to identify components that must be recycled.
(The console is shown with the chassis cover removed for clarity.)