4.Splice programAttenuation splice mode
4. Splice program
5: Select “ATTN Coef.” to edit
the value.
6: Enter the value of ATTN Coef.
If the value is increased, fiber
offset will become large and a
splice loss will be higher. If the
value is decreased, a splice loss
will be lower.
7: Tap Set icon ( ) to perform a
splice.
Check an actual splice loss. If the
actual splice loss still does not
match the target splice loss, go
back to step 5~6 to adjust ATTN
Coef again.
Item Details
Operator Administrator
View Edit View Edit
Target Loss Sets the target splice loss for attenuation.
Fiber Core
Offset for ATTN
Sets the target core offset value for attenuation.
This parameter will be changed in association
with “Target loss”.
ATTN Coef.
This parameter is the attenuation splice coeffi-
cient. If actual splice loss measured does not
match “Target Loss”, this parameter can be
used to adjust.