EasyManua.ls Logo

Sun Microsystems DGA 5000 - Zero-Calibration

Sun Microsystems DGA 5000
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.2.4
Zero-calibration
Zero-calibration, which allows the Gas Module to determine the correct
zero reference and gain settings.
During this time:
-
The zero-calibration pop-up screen will be displayed showing the
elapsed calibration time.
An zero-calibration may be triggered in three ways:
-
As part of the start-up procedure or on return from standby mode
(initial zero-calibration).
-
Automatically by the Gas Bench as required (automatic zero-
calibration; see note below).
-
Manually by the operator from Operation - The Free
Measurement Procedure (see 8) (manual zero-calibration).
An initial or manual zero-calibration (but not an automatic zero-
calibration) will be followed by a HC residue check (see below).
If the system is configured to require a Daily Leak Check, the unit will
prompt the operator to perform this at the end of the zero-calibration
period. Until a successful leak check has been performed all
measurements will be blocked and it will not be possible to proceed
beyond the Gas Analyser Menu (see 6.3: „Daily Leak Check‟ below
for further details).
Note:
The gas bench automatically assesses the stability of the zero reference
and gain settings and will perform an zero-calibration as and when
required. The frequency with which an automatic zero-calibration is
performed will depend on the stability of the environment in which the
analyser is operated. If a CO2 concentration of 0.5 % vol. or greater is
detected at the time the gas bench determines that an automatic zero-
calibration is necessary, the software will assume that gas
measurements are currently being taken. In this case the zero-
calibration will be delayed until the CO2 measurement has fallen below
0.5% vol. or for a maximum of 30 minutes. The Error/Warning toolbar
button will be displayed. If this button is pressed, error/warning code 06
will be seen, advising that a zero-calibration is pending
29
C H A P T E R 6

Table of Contents

Other manuals for Sun Microsystems DGA 5000