Note:
The three statuses will be run in sequence upon power-up or when
returning from standby mode.
a.
The Zero-calibration and HC residue check will be run in
sequence following the manual initiation of an auto-calibration.
b.
An automatically triggered auto-calibration will not be followed
by an HC residue check.
c.
In the event of maintenance function (e.g. leak check) being
selected whilst a power-up or manually initiated auto-calibration is
in progress, the auto-calibration pop-up window will be displayed
and the auto-calibration will be completed. The subsequent HC
residue check will be postponed or aborted until the maintenance
mode is exited.
9.8 The Error/Warning Button
The “Error/warning” button will appear on the right hand side of the
toolbar whenever an error or warning condition is detected. Examples of
such conditions are messages concerning the status of the O2 Sensor,
insufficient gas flow, daily leak check requirement etc.
•
Select the Error/Warning button in order to review the error/warning
messages.
The Error/Warning pop-up window will be displayed showing the
error/warning code together with a brief description of the cause of
the error/warning message.
The Error/Warning pop-up window may be closed at any time by
selecting the “Close” button.
The Error/Warning button will disappear automatically when the
conditions giving rise to the error/warning message(s) are rectified.
Error/Warning codes and messages may be of two types:
Messages relating to the DGA 5000 system.
-
Messages relating to the gas bench
The possible error codes are show by the tables below.
68