EasyManua.ls Logo

Sun Microsystems DGA 5000 - 9 Maintenance; 9.1 System Checks; 9.2 Leak Check (Vacuum)

Sun Microsystems DGA 5000
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The “Maintenance” section of this manual is divided into three main sub-
sections:
-
Operator initiated system checks.
-
Routine maintenance procedures.
-
Error codes and warnings.
9.1 System Checks
The operator may initiate the following system checks:
-
Leak Check (Vacuum).
-
Leak Check (Gas).
-
Gas Calibration Check
-
Check and/or Install the O2 Cell.
All the above checks are initiated via the Gas Analyser Maintenance
Menu (see 5.3).
9.2 Leak Check (Vacuum)
Note:
A daily check of the gas analyser pneumatic circuit for air leaks that
could compromise the accuracy of the test results is a mandatory
requirement of many country specific test procedures (see 6.3: „Daily
Leak Check‟). If the DGA 5000 is configured for an approval type that
requires such a daily leak check, the appropriate settings will be made at
the time of the installation of the test software. In this case the unit will
not commence normal operation unless a satisfactory leak check has
been carried out. In order to ensure the continued accuracy of the test
results, it is recommended that a daily leak check be carried out at the
start of the day‟s operation even in cases where such a test is not a
statutory requirement. To conduct a Leak Check (Vacuum):
Select the “Leak Check (Vacuum)” option from the Gas Analyser
Maintenance Menu (see 5.3)
52
9
Maintenance

Table of Contents

Other manuals for Sun Microsystems DGA 5000