Figure 9-10 O2 Sensor Replacement - Auto-calibration
The unit will initiate the auto-calibration and the auto-calibration pop-up
window will appear showing the auto-calibration countdown. At the end
of the auto-calibration the screen will show the message “Installation
complete” and prompt the operator to press “Exit” to leave the
installation procedure.
Press “Exit” to return to the Gas Analyser Maintenance screen.
Replace the O2/NO sensor cover plate (if applicable).
62
CHECK AND/OR INSTALL THE O2 CELL