EasyManuals Logo

Supmeter BST100-B11 User Manual

Supmeter BST100-B11
73 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #43 background imageLoading...
Page #43 background image
42
6.3.7 Load Parameters (LoAd)
No.
Sign
Parameter
Range
Default
Set
P500
-dot
Load Decimal Point
0: o
1: o.o
2: o.oo
1
P501
rAng
Load Range
0.01~5000.00kg/m
([P107] Scale Unit = kg)
100.00
0.01~5000.00g/m
([P107] Scale Unit = g)
P502
-HI-
Load Upper Limit
(0.0~100.0)% Load Range
100.0%
P503
-Lo-
Load Lower Limit
(0.0~100.0)% Load Range
0.0%
P504
FILt
Load Filter
1~20
10
P505
LdCA
Speed for Calculation
0.000 (Use external speed for Load
calculation)
0.001~5.000m/s (Used for Load
calculation)
0.000
6.3.8 Weight Record Parameters (SAVE)
No.
Sign
Parameter
Range
Default
Set
P600
SHFn
Shifts per Day
1: One Shift; 2: Two Shifts;
3: Three Shifts; 4: Four Shifts
3
P601
SHF1
Shift1 Time
00:00~23:59
07:59
P602
SHF2
Shift2 Time
00:00~23:59
15:59
P603
SHF3
Shift3 Time
00:00~23:59
23:59
P604
SHF4
Shift4 Time
00:00~23:59
23:59
P605
APrt
Auto-print
0: OFF; 1: Per Hour
2: Per Shift; 3: Per Day
0
P606
ACLn
Auto Clear Screen
0: OFF; 1: Per Shift
2: Per Day; 3: Per Month
„Auto-clear has no effect on recording of
weight per shift.
0
www.zemicusa.info

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Supmeter BST100-B11 and is the answer not in the manual?

Supmeter BST100-B11 Specifications

General IconGeneral
BrandSupmeter
ModelBST100-B11
CategoryController
LanguageEnglish

Related product manuals