AMI Codes-II
Operation
A-96.250.571 / 050517 39
5.2. Display
Relay status, symbols
A RUN normal operation
HOLD input closed or cal delay: Instrument on hold (shows
status of signal outputs).
OFF input closed: control/limit is interrupted (shows status
of signal outputs).
B ERROR Error Fatal Error
C Reagent low, indicates remaining reagents in% (17% = 340 ml)
D Keys locked, transmitter control via Profibus
E Time
F Process values
G Sample Temperature
H Cleaning solution low, indicates remaining cleaning solution in%
I Sample flow in B/s
K Relay status
upper/lower limit not yet reached
upper/lower limit reached
control upw./downw. no action
control upw./downw. active, dark bar indicates control intensity
motor valve closed
motor valve: open, dark bar indicates approx. position
timer
timer: timing active (hand rotating)
RUN
23 °C
23 B/s
15:20:18
R1
R2
ppm
0.15
AB C D
F
E
G
H
K
I

Table of Contents