48 A-96.250.571 / 050517
AMI Codes-II
Maintenance
6.3.2 Reagents for Measuring Monochloramine and Ozone
2 L of reagents last for 15 days with 2 min. interval:
1 x 50 ml of Oxycon On-line DPD
1 x 240 gr of Oxycon On-line Buffer
1 x 60 gr of Reagent Oxycon On-line KI
4 liters of demineralized water
Prepare
Oxycon
On-line DPD
1 Rinse the canister [E] labelled OXYCON ON LINE DPD-Re-
agent with demineralized water.
2 Fill the canister up to the 2 liter mark with demineralized water.
3 Slowly pour the content of a bottle of concentrate Oxycon On-
line DPD into the canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well..
6 Put the canister [E] into the holder [G].
7 Remove the screw cover and insert the suction lance [A] and
tighten the screw cover [B].
Prepare
Oxycon
On-line Buffer
& KI
1 Rinse the canister [F] labelled “OXYCON ON LINE Buffer” solu-
tion with demineralized water.
2 Fill the canister up to the 2 liter mark with demineralized water.
3 Slowly pour the content of one bag of Oxycon On-line Buffer
into the canister.
Avoid splashing!
4 Add the content of one bag Oxycon On-line KI to the same can-
ister.
5 Close the canister with the screw cover and tighten it well.
6 Mix the demineralized water and the reagents well.
7 Put the canister [F] into the holder [G].
8 Remove the screw cover and insert the suction lance [A] and
tighten the screw cover [B].
9 Fill reagent system. See Fill or Flush Reagent System, p. 36.