EasyManua.ls Logo

Swann AMI Codes-II - Instrument Specification

Swann AMI Codes-II
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AMI Codes-II
Product Description
A-96.250.571 / 050517 13
2.1. Instrument Specification
Power Supply Voltage:
Power consumption:
100240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
max. 30 VA
Electronics Aluminium with a protection degree of IP 66 / NEMA 4X
housing Ambient temperature:
Limit range of operation:
Storage and transport:
Humidity:
Display:
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C
1090% rel., non condensing
backlit LCD, 75 x 45 mm
Sample
requirements
Flow rate:
Temperature:
Inlet pressure:
Outlet pressure:
min. 10 l/h
550 °C (41122 °F)
0.152 bar (228 PSI)
pressure free
NOTICE: No oil, no grease, no sand.
On-site The analyzer site must permit connections to:
requirements Sample inlet:
Sample outlet:
Tube 6 x 8 mm
1/2” hose nozzle for flexible tube
diam. 20x15 mm
Measuring
Range
Ozone
0.0001.000 ppm
Accuracy
± 0.005 ppm
HOCL, free chlorine, monochloramine
0.001.00 ppm
1.003.00 ppm
3.005.00 ppm
± 0.01 ppm
± 0.06 ppm
± 0.2 ppm
Chlorine dioxide, iodine, bromine
0.002.00 ppm
2.006.00 ppm
± 0.02 ppm
± 0.12 ppm

Table of Contents

Related product manuals