AMI Codes-II
Maintenance
A-96.250.571 / 050517 43
6.2. Stop of Operation for Maintenance
1 Put the suction lances into a bucket with clean water.
2 Start <Fill system>.
The reagent tubes are flushed with water.
3 Remove the suction lances from the water.
4 Start <Fill system> again.
The water will be pumped out of the reagent tubes.
5 Stop sample flow.
6 Wait until level in flow cell is empty.
7 Shut off power of the instrument.
If pH option is installed:
8 Remove the pH sensor [A] from the flow cell.
9 Fill the protective cap [B] with water.
10 Put the protective cap onto the sensor tip.
A
B
pH sensor
Protective cap
A
B