50 A-96.250.571 / 050517
AMI Codes-II
Maintenance
6.5. Calibration
Process
Calibration of
DIS
NOTICE: Perform process calibration for free chlorine / total
residual chlorine only if:
the sample concentration is close to the desired process
value (stable value)
you are sure that the reagents are mixed completely and
correctly
if the difference to the manual measurement is significant.
Keep in mind the accuracy of your manual measurement.
Use CHEMATEST 25 (or equivalent photometer) to determine the
sample disinfectant concentration. Determine the sample disinfec-
tant value by 3 manual DPD measurements. Calculate the average
value. Compare this value to the value, indicated by the AMI.
Enter process value (mg/l = ppm) under menu 3.1.1, p. 78 for Pro-
cess DIS
Possible error message see Calibration Errors, p. 65.
Zero A zero is automatically done before each measurement.
Press 3 x [Exit]
3.1.1
Process pH
Process DIS.
Calibration
Standard pH
Enter
3.1.1.4
Process DIS.
Current Value x.xx ppm
Slope x.xxx
Save <Enter>
Process Value x.xx ppm
3.1.1.4
Process DIS.
Current Value x.xx ppm
Slope x.xxx
Save <Enter>
Process Value x.xx ppm
3.1.1.5
Process DIS.
Current Value x.xx ppm
Slope x.xxx
Calibration Successful