46 A-96.250.571 / 050517
AMI Codes-II
Maintenance
Canister set up
6.3.1 Reagents for Free Chlorine, Chlorine Dioxide, Bromine
and Iodine
To prepare 2 L of reagents the following chemicals are necessary:
1 x 50 ml of Oxycon On-line DPD
1 x 240 gr of Oxycon On-line Buffer
4 liters of demineralized water
Prepare
Oxycon
On-line DPD
1 Rinse the canister [E] labelled “OXYCON ON LINE DPD-Re-
agent” with demineralized water.
2 Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3 Slowly pour the content of one bottle of concentrate Oxycon
On-line DPD (50 ml) into the canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
A
B
C
D
E
F
G
Suction lance without
level detector
(canister 1)
Suction lance with
level detector
(canister 2)
Level detector
2 L mark
Canister 1:
Oxycon on-line DPD
Canister 2:
Oxycon on-line
Buffer
Holder
OXYCON ON-LINE
DPD-Reagenz
DPD Réactif
DPD Reagent
2L
OXYCON ON-LINE
Pufferlösung
Tam p on
Buffer Solution
2L
B
C
E
D
G
F