AMI Codes-II
Maintenance
A-96.250.571 / 050517 47
6 Put the canister [E] into the holder [G].
7 Remove the screw cover, insert the suction lance [A] and tight-
en the screw cover.
Prepare
Oxycon
On-line Buffer
1 Rinse the canister [F] labelled “OXYCON ON LINE Buffer” solu-
tion with demineralized water.
2 Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3 Slowly pour the content of one bag of buffer Oxycon On-line
(240 gr) into the canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
6 Put canister [F] into holder [G].
7 Remove the screw cover and insert the suction lance [B] and
tighten the screw cover.
8 Fill reagent system. See Fill or Flush Reagent System, p. 36.