14 A-96.250.111 / 211019
AMI Trides
Product Description
2.2. Instrument Specification
Power Supply Voltage: 100240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
Power consumption: max. 30 VA
Sample
requirements
Water consumption: aprox. 40 l/h
Temperature: 545 °C (41113 F)
Inlet pressure: 0.152 bar (2.229 PSI)
Outlet pressure: pressure free
Min. sample conductivity: 5 S/cm
no oil and no grease
On-site
requirements
The analyzer site must permit connections to:
Sample inlet: Hose nozzle R ¼” x 6 mm
Sample outlet: G ½” adapter for flexible tube
15 x 20 mm
Electronics
housing
Aluminium with a protection degree of IP 66 / NEMA 4X
Ambient temperature: - 10 to + 50 °C
Humidity: 1090% rel., non condensing
Display: backlit LCD, 75 x 45 mm