AMI Trides
Program List and Explanations
A-96.250.111 / 211019 67
9. Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5 Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
1.2 Message List
1.2.1 Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Desig.: Designation of the instrument.
Version: Firmware of instrument (e.g. V6.20-06/16)
2.1.3 Factory Test: Test date of the Instrument and Motherboard
2.1.4 Operating Time: Years / Days / Hours / Minutes / Seconds
2.2 Sensors
2.2.1 Trides Sensor:
Current value: Shows the current disinfectant measuring value in
ppm
Raw value 1: Shows the actual sensor current in µA, without
compensation of temperature (and pH).
Raw value 2: Shows the actual sensor current in µA, with
compensation of temperature (and pH).
Ref. voltage: Shows the actual voltage of counter electrode (CE) in
mV. Values for most applications are within 800 - 1300 mV.
2.2.1.5 Cal. History: Review the diagnostic values of the last calibrations.
Number, Date, Time, Offset [µA], Slope [µA]