6.5. Cleaning of pH Electrode
Clean
pH Electrode
1 Pull the pH electrode [A] out of the constant head.
2 If necessary wipe the electrode shaft and the green tip cautious-
ly with a soft, clean, and damp paper tissue.
3 Remove grease with a tissue moistened with alcohol.
4 If the electrode is very dirty, put its tip into 1% diluted hydrochlo-
ric acid for roughly 1 min.
5 Afterwards rinse the electrode tip thoroughly with clean water.
6 Install the electrode into the constant head again.
7 Let the electrode run-in for 1 h before the first calibration.