AMI Trides
Maintenance
A-96.250.111 / 211019 43
6.2. Cleaning of Trides protective filter
If the protection filter shows deposits, proceed as follows:
1 Close the flow regulating valve [A].
2 Close the sample main tap before the filter.
3 Unscrew and remove the filter vessel [C] from the flow cell
block.
4 Unscrew and remove the filter from [B] the flow cell block.
5 Backwash the filter under pressure of tap water. Clean the out-
side of the filter.
6 Install the filter and filter vessel again.
7 Open sample supply and flow regulating valve again.
A
B
C
Flow regulating valve
Filter
Filter vessel