AMI Trides
Troubleshooting
A-96.250.111 / 211019 61
E014 Case Temp. low
– check case/environment temperature
– check programmed value, see 5.3.1.6,
p. 78
E015 TRIDES Reference
– Check if conductivity value of sample is
> 5 µS/cm.
– Calculate: actual Trides signal /
average value 3 DPD meas. = must be
>2 µA/ppm. If not, clean Trides sensor.
– Exchange reference electrode.
E017 Control Timeout
– check control device or programming in
Installation, Relay contact, Relay 1and 2,
5.3.2 and 5.3.3, p. 78
E024 Input active
– See If Fault Yes is programmed in
Menu see 5.3.4, p. 83
E026 IC LM75
– call service
E028 Signal output open
– check wiring on signal outputs 1 and 2
E030 EEProm Frontend
– call service
E031 Cal. Recout
– call service
E032 Wrong Frontend
– call service
E033 Power-on
– none, normal status
E034 Power-down
– none, normal status
Error Description Corrective action