5.3.1.2.25 Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E004 is displayed in the message
list.
Range: -3.00 to + 15.00 pH
5.3.1.2.35 Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range. 0 –2 pH
5.3.1.2.45 Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec
5.3.1.3 Sample Flow: Define at which sample flow an alarm should be is-
sued.
5.3.1.3.1 Flow Alarm: Program if the alarm relay should be activated if there
is a flow alarm. Choose between yes or no. The flow alarm will
always be indicated in the display, pending error list, saved in the
message list and the logger.
Available values: Yes or no
NOTICE: Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.3.2 Alarm High: If the measuring values rises above the programmed
value E009 will be issued.
Range: 400–9000 rpm
5.3.1.3.35 Alarm Low: If the measuring values falls below the programmed
value E010 will be issued.
Range: 200–350 rpm
5.3.1.3.45 Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec