EasyManua.ls Logo

TA Instruments TGA 2950 - Page 174

TA Instruments TGA 2950
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix B
B–36 TA I
NSTRUMENTS
TGA 2950
set point, and not so high as to cause percent/
minute overshoot, transitions will appear largely
the same from one heating rate to the next. A
word of caution, however: always keep in mind
that the background (baseline) rate of weight
change will be accelerated at higher maximum
heating rates. This will move the background
percent/minute closer to the control point you
have selected with the resolution setting. If
noise is present in the background percent/
minute, it may fool the control algorithm into
thinking a transition is starting and result in a
premature reduction in heating rate to avoid
predicted set point overshoot. In extreme cases,
heating rate cycling can occur, particularly at
high sensitivity settings. (See Figure B.13 on
page B-61) To cure this problem reduce maxi-
mum heating rate and/or sensitivity setting.

Table of Contents

Related product manuals