Introducing the TGA 2950
TA I
NSTRUMENTS
TGA 29501–8
Table 1.1
TGA 2950 Keypad
Function Keys
(continued)
Key/Function Explanation
STOP If an experiment is
running, this key endsthe
method normally, as
though it had run to
completion; i.e., the
method-end conditions
go into effect and the data
that has been generated is
saved. This is the same
function as Stop on the
controller.
If an experiment is not
running (the instrument is
in a stand-by or method-
end state), the STOP key
will halt any activity (air
cool, all mechanical
motion, etc.).
REJECT If an experiment is
running, SCROLL-STOP
ends the method normal-
(Hold down ly, as though it had run to
SCROLL and completion; i.e., the
press STOP) method-end conditions go
into effect and the data
that has been generated is
discarded. This is the
same function as Reject
on the controller.
(table continued)