Taiwan Biophotonic Corporation
53
1 Press SWITCH button to select:
“✓(in GREEN)”: restore all settings to factory
default value.
“X (in RED)”: about the reset process.
2 Press ENTER button to accept the selection and
return to “SETTING” menu.
SpO
2
/PR high (or low) limit
1 Press SWITCH button to select the target setting for
the SpO
2
high limit, and press ENTER button to
accept the setting and move to SpO
2
low limit
setting.
2 Repeat the same process above for SpO
2
low limit,
PR high limit and PR low limit.
3 Once the PR low limit is set, press ENTER button to
accept the settings and return to “SETTING” menu.
NOTE:
The table below shows the range and default value of the
PR/SpO
2
limits:
86~100%, with 1% step size
85~95%, with 1% step size
75~250bpm, with 5bpm step
size