53
1 Press SWITCH button to select:
(in GREEN): restore all settings to factory
default value.
X (in RED): about the reset process.
2 Press ENTER button to accept the selection and
return to SETTING menu.
SpO
2
/PR high (or low) limit
1 Press SWITCH button to select the target setting for
the SpO
2
high limit, and press ENTER button to
accept the setting and move to SpO
2
low limit
setting.
2 Repeat the same process above for SpO
2
low limit,
PR high limit and PR low limit.
3 Once the PR low limit is set, press ENTER button to
accept the settings and return to SETTING menu.
NOTE:
The table below shows the range and default value of the
PR/SpO
2
limits:
Parameter
Ranges
Defaults
SpO
2
High Limit
86~100%, with 1% step size
Off
SpO
2
Low Limit
85~95%, with 1% step size
85%
PR High Limit
75~250bpm, with 5bpm step
size
170bpm